In LC-MS/MS and GC-MS/MS using triple quadrupole ($QqQ$) mass spectrometers, MRM stands for Multiple Reaction Monitoring.
In MRM, the first quadrupole ($Q_1$) selects a specific precursor ion, the collision cell ($q_2$) fragments it using collision-induced dissociation (CID), and the third quadrupole ($Q_3$) monitors specific characteristic product ions, providing extraordinary selectivity and sensitivity for trace residue quantitation.
Conclusion: MRM stands for Multiple Reaction Monitoring (Option 2).
Correct option: (2) Multiple reaction monitoring
Explanation not matching this question?क्या व्याख्या इस प्रश्न से मेल नहीं खा रही?Report
Q.7
Agricultural Chemicals- ICAR JRF/SRF 2024 (Q.108)
MCQ (Single Correct)+4−1
In NMR, protons that are involved in hydrogen bonding (i.¢.-OH or -NH) are usually observed over a wide range of chemical
shifts. This is due to the deshielding that occurs in the hydrogen bond. Since hydrogen bonds are dynamic, constantly
forming, breaking and forming again, there will be a wide range of hydrogen bonds strengths and consequently a wide range
of deshielding. Thus experimentally -OH and -NH can be identified by carrying out
Because deuterium ($^2\text{H}$) resonates at a completely different frequency, the active $-OH$ or $-NH$ resonance peak disappears from the $^1\text{H}$ spectrum (collapsing into the $ ext{HOD}$ peak at $\delta \approx 4.7\text{ ppm}$), conclusively identifying exchangeable protons.
Conclusion: Identified via $\text{D}_2\text{O}$ exchange experiment (Option 2).
Correct option: (2) D20 exchange
Explanation not matching this question?क्या व्याख्या इस प्रश्न से मेल नहीं खा रही?Report
Q.8
Agricultural Chemicals- ICAR JRF/SRF 2024 (Q.109)
MCQ (Single Correct)+4−1
Orbitrap technology is mainly used for the determination purpose of
The Orbitrap mass analyzer (invented by Alexander Makarov) captures ions in electrostatic oscillation around a central spindle electrode, determining $m/z$ from the harmonic axial oscillation frequency.
It is globally celebrated in metabolomics and pesticide residue analysis for providing ultra-high mass resolving power ($\gt 100,000$ to $500,000$) and sub-ppm accurate mass measurement, enabling unambiguous elemental formula determination of unknown analytes.
Conclusion: Orbitrap is primarily used for Accurate Mass determination (Option 1).
Correct option: (1) Accurate mass
Explanation not matching this question?क्या व्याख्या इस प्रश्न से मेल नहीं खा रही?Report
Report Issue / Suggest Correctionत्रुटि रिपोर्ट / संशोधन सुझाव
No official replies published yet for this questionइस प्रश्न पर अभी कोई आधिकारिक समाधान प्रकाशित नहीं है
Have an objection, typo, or doubt? Submit a report in Tab 1. Our subject experts will audit the question and publish the official resolution here.यदि आपको इस प्रश्न या उत्तर कुंजी में कोई आपत्ति है, तो 'त्रुटि रिपोर्ट' टैब से विवरण भेजें। हमारी विषय विशेषज्ञ टीम समीक्षा कर आधिकारिक समाधान प्रकाशित करेगी।