Hounsfield unit is a term, associated with which of the following imaging technique?
हाउन्सफील्ड यूनिट एक शब्द है, जो निम्नलिखित में से किस इमेजिंग तकनीक से संबंधित है?
Diagnostic imaging principles of Computed Tomography (CT):
- Hounsfield Unit (HU): Named after Sir Godfrey Hounsfield (co-inventor of CT), HU is a quantitative scale describing radiodensity of tissues in CT scanning relative to distilled water and air.
- Scale calibration: Distilled water is defined as $0\text{ HU}$, air is $-1000\text{ HU}$, compact cortical bone is $+1000\text{ to } +3000\text{ HU}$, and soft tissues range from $+20\text{ to } +70\text{ HU}$.
- MRI utilizes T1/T2 relaxation times; ultrasonography measures acoustic impedance / echogenicity; nuclear scintigraphy measures radioactive gamma counts.
Option analysis: - MRI uses relaxation times (T1, T2) and signal intensity.
- (CT scan): Correct. Hounsfield units quantify attenuation values in computed tomography.
- Ultrasound uses echogenicity and decibels.
- Scintigraphy measures radioactive disintegration rates.
कंप्यूटेड टोमोग्राफी (सीटी स्कैन) में हाउन्सफील्ड यूनिट:
- हाउन्सफील्ड यूनिट (HU): सीटी स्कैन के सह-आविष्कारक सर गॉडफ्रे हाउन्सफील्ड के नाम पर रखी गई यह इकाई ऊतकों के एक्स-रे अवशोषण या रेडियोडेंसिटी को मापती है।
- पैमाना: आसुत जल का मान $0\text{ HU}$, हवा का मान $-1000\text{ HU}$, सघन अस्थि का मान $+1000$ से $+3000\text{ HU}$ तथा कोमल ऊतकों का मान $+20$ से $+70\text{ HU}$ निर्धारित किया गया है।
- एमआरआई में T1/T2 रिलैक्सेशन समय, अल्ट्रासाउंड में प्रतिध्वनि (इकोजेनिसिटी) तथा सिंटिग्राफी में गामा विकिरण गिना जाता है।
विकल्प विश्लेषण: - एमआरआई सिग्नल तीव्रता पर आधारित है।
- (सीटी स्कैन): सही। सीटी स्कैन में ऊतकों के घनत्व की प्रामाणिक इकाई।
- ध्वनि तरंगों पर आधारित है।
- रेडियोधर्मी विकिरण पर आधारित है।